AAAAA级少妇高潮大片免费看,JIZZ在线观看免费网站,香港三级日本三级妇三级,女学生处破外女出血视频

產品展示
首頁 > 產品展示 > > 其他檢測儀器 > CS600高精度碳硫測定儀

CS600高精度碳硫測定儀

簡要描述:簡單說明:CS600提供了一系列*的解決方案檢測碳和硫。當今的燃燒和紅外相結合的檢測技術使您能更準確、精確、使用方便靈活的剖析各種類型的金屬和無機材料。

  • 產品型號:CS600
  • 廠商性質:生產廠家
  • 更新時間:2024-10-08
  • 訪  問  量:1802
詳情介紹

詳細說明:

CS600提供了一系列*的解決方案檢測碳和硫。當今的燃燒和紅外相結合的檢測技術使您能更準確、精確、使用方便靈活的剖析各種類型的金屬和無機材料。

Carbon and Sulfur in Steel, Nickel-Base, and Cobalt-Base Alloys
 
Instrument
CS600-Series
 
Sampling and Sample Preparation
Surface contamination on the sample can cause significant errors in the analytical data; therefore, care must be taken to ensure a clean, representative sample is analyzed. Solid
samples should be abraded with a clean file, rinsed in acetone and dried with warm air prior to analysis. Samples that cannot be abraded due to irregular shapes should be rinsed in a
suitable solvent such as acetone, and dried with warm air. Care
must be taken to remove all traces of the solvent. If a sample is porous, refrain from using solvents,
as it will be difficult to remove all traces of the solvent by drying. Refer to ASTM E1806 for additional sampling and sample preparation information.
 
Accessories
528-018 or 528-018HP Ceramic Crucibles (preheated)*; LECOCEL II (501-008) or
LECOCEL II HP (502-173) accelerator.
 
*Ceramic crucibles are baked in a muffle or tube furnace (LECO TF-10) at 1250oC for a minimum of
15 minutes, or at 1000oC for 40 minutes. The crucibles are removed from the furnace, allowed to cool for

 
 
 
 
1 to 2 minutes, and are transferred to a desiccator for storage. If the crucibles are not used within four hours, they should be re-baked.

 
Calibration
There are several suitable calibration samples available from LECO. Likewise, NIST, JK, JSS, and BCS are certified bodies that have a variety of certified reference materials (SRM/CRM) available as well. It is important that both high- and low-carbon ranges are calibrated. Single or multipoint calibration curves
can be utilized. Refer to the operator's instruction manual for details.
 
Method Parameters
Purge Time (seconds):
1**
 
Delay Time (seconds):
8**
 
Furnace Low Power (%):
100
 
Furnace High Power (%):
100
 
Furnace Ramp Rate:
0
 
 
 
Carbon
 
Sulfur
Minimum Timeout (seconds):
40
40
Comparator Level:
1.00
1.00
Significant Digits:
4 or 5
4 or 5
Integration Delay:
0
0
 
 


 

**For optimum analytical performance on samples with low carbon and
sulfur content 0.05%), a purge time of 10 seconds and a delay time of
20 seconds is recommended.

 
 
CS600


 

Procedure
1.   Prepare instrument for operation as outlined in the operator's instruction manual.
2.   Determine blank.
a.     Enter 1.0000 g weight into weight stack. b.      Add ~1.5 g of accelerator to crucible.
c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
d.     Repeat steps 2a through 2c a minimum of five times.
e.     Enter blank following procedure outlined in the operator's instruction manual.
3.   Calibrate.
a.     Weigh ~1.0 g calibration sample into crucible and enter weight into weight stack.
b.     Add ~1.5 g of accelerator on top of sample.
c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
d.     Repeat steps 3a through 3c a minimum of five times for each calibration sample intended for calibration.
e.     Calibrate using the procedure outlined in the operator's instruction manual.
4.   Analyze Samples.
a.     Weigh ~1.0 g sample into crucible and enter weight into weight stack.
b.     Add ~1.5 g of accelerator on top of sample.
c.      Place crucible on furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
 
Note:   Some "solid" samples, notably nickel and/or cobalt-base alloys, are difficult to combust and may require additional accelerator. Likewise, reducing the sample to small chunks, chips, or pieces will facilitate combustion. For additional information on Carbon and Sulfur Determination using an Induction Furnace IR detection instrument, request LECO's Application Tips CD (209-050-055).
 
Typical Results

Sample
Weight (g)
C %
S %
Sample
Weight (g)
C %
S %
LECO
1.0
0.605
0.0176
LECO
1.0
0.00089
0.00062
501-505
 
0.605
0.0172
502-348
 
0.00089
0.00067
Steel Ring
 
0.605
0.0169
Low C and S
 
0.00094
0.00068
0.607 % C
 
0.607
0.0171
Steel Pin
 
0.00095
0.00060
0.0172% S
 
0.607
0.0174
0.0009 % C
 
0.00085
0.00064
 
 
0.605
0.0176
0.0007 % S
 
0.00085
0.00061
 
 
0.606
0.0178
 
 
0.00090
0.00062
 
 
0.604
0.0173
 
 
0.00090
0.00060
 
 
0.606
0.0173
 
 
0.00090
0.00065
 
 
0.605
0.0168
 
 
0.00089
0.00055
 
X =
0.606
0.0173
 
X =
0.00090
0.00062
 
s =
0.001
0.0003
 
s =
0.00003
0.00004
 
LECO
 
1.0
 
0.0375
 
0.0223
 
LECO
 
1.0
 
0.0414
 
0.0025
501-501
 
0.0380
0.0221
501-933
 
0.0414
0.0027
Steel Ring
 
0.0375
0.0222
Ni-Base Alloy
 
0.0413
0.0027
0.0377 % C
 
0.0377
0.0222
Chip Sample
 
0.0414
0.0028
0.0220 % S
 
0.0376
0.0223
0.042 % C
 
0.0415
0.0027
 
 
0.0376
0.0222
0.003 % S
 
0.0413
0.0028
 
 
0.0373
0.0221
 
 
0.0414
0.0028
 
 
0.0376
0.0222
 
 
0.0415
0.0028
 
 
0.0377
0.0223
 
 
0.0412
0.0027
 
 
0.0377
0.0222
 
 
0.0414
0.0029
 
X =
0.0376
0.0222
 
X =
0.0414
0.0027
 
s =
0.0002
0.0001
 
s =
0.0001
0.0001


 

留言框

  • 產品:

  • 您的單位:

  • 您的姓名:

  • 聯系電話:

  • 常用郵箱:

  • 省份:

  • 詳細地址:

  • 補充說明:

  • 驗證碼:

    請輸入計算結果(填寫阿拉伯數字),如:三加四=7
技術支持:化工儀器網   sitemap.xml   管理登陸
©2024 版權所有:寧波市鄞州瑾瑞儀器設備有限公司   備案號:浙ICP備14028217號-3

浙公網安備33020402000311號